THERMO-SCIENTIFIC ORION PHOSPHATE ANALYZER (2295)
PURPOSE : The purpose of this document is to make understand the standard calibration and maintenance procedure of THERMO-SCIENTIFIC ORION PHOSPHATE Analyzer (2295) and according to this procedure, PHOSPHATE Analyzer will be calibrated as per the OEM recommendations for calibration and can help to maintain accuracy and reliability of readings.
SOURCE: USER MANUAL
GENERAL RESPONSIBILITIES:
· Ensure
that the Reagents Solutions are made as per the OEM. Reagents must be replaced
at the time of calibration and confirm with chemist that all the solutions are
mixed thoroughly before use.
· Check
for leaks, particularly around the sample and drain pipe work connections.
· Confirm
sample flow by checking delivery to the constant-head unit and effluent from
the drain.
· Check
liquid levels in the reagent.
· Inspect
all tubing and liquid handling components for leaks and deterioration.
· Check
for malfunction indications on the analyzer display.
· Ensure
following PPE as per EHS procedure no. xxxxxxxxxxxxxxxxx Safety shoe, Helmet,
Dust Respirator, and Protective Eye Wear and proper hand Wear.
· Ensure Healthiness & Safety of the entire activities with respect electrical and Mechanical system.
Job Specific Responsibilities:
Sl.No |
Particulars |
Responsibility |
1 |
Issue and
Closure of work permit |
Engineer
in charge/ Operation unit & shift in charge |
2 |
Ensuring isolation |
Operation
shift/unit in charge |
3 |
Monitoring availability of required tools & tackles and
implementation of relevant OH&S and Environmental Procedures |
Engineer
in charge/Supervisors |
4 |
Checking the Quality of Job and Providing feedback to ensure task
Completion as per the procedure |
Engineer in charge |
5 |
Performing of tasks in compliance with Instructions and safe working
with Equipment |
Technicians |
6 |
Carrying out relevant monitoring tasks such measurement, verification
and validation to ascertain completion of Scheduled tasks. |
Engineer in charge |
STARTUP
PROCEDURE:
·
Power
on instrument.
· To allow longer lifetime of unit, the pump-head tubing are not fully engaged when the unit is shipped. Refer to Plumbing Connections Section.
·
Reset
tubes in pumps heads.
Note: This step must be repeated for all four pump
heads.
i.
Remove
tape.
ii. Slide tube retainer into place along keyway.
iii. Install tubing cover by squeezing both sides,
and sliding into grooves.
Install Reagents. Very carefully match reagent
numbers and colors in their proper order.
Prime the tubing. Please repeat the following
sequence 3 times.
a. Press Menu.
b. Select Services.
c. Select Prime Fluidics.
d. Select Prime All.
Note: This
procedure will run automatically and take approximately 10 minutes per cycle.
·
Once the above 4 steps have been completed,
(4. a, b, c, d) press Stop, Select Run. Allow the unit to operate overnight or
for at least 12 hours, to become stabilized in the local environment.
GENERAL INSTRUCTION FOR USER INTERFACE COMPLIANCE:
·
The
user-interface consists of a keypad and a graphical LCD display. The keypad is
located to the right of the display. The "F1-F4" function keys
(soft-keys) take on different meanings depending on the menu selection.
·
The
column on the left of the soft-keys labels the keys with their appropriate
functions.
·
The
RUN key starts the analysis and the STOP key stops the current functioning of
the
Instrument.
·
The +
and - keys allow for navigation between pages.
·
The
ENTER key confirms selections made in the menus. Additional navigation is
provided by the soft-keys (in appropriate menus), for left / right or up / down
cursor movement.
·
The
graphical LCD display shows the date, analysis label, time, navigational menus,
information and help to the user, current user, instrument-working mode,
current status of the instrument and the soft-key labels.
· The MENU key is used to access the menus.
Display Fields:
- DATE FIELD shows the current date.
- CURSOR indicates current selection.
- INSTRUMENT NAME FIELD shows instrument nametag.
- TIME FIELD shows present time.
- DISPLAY FIELD shows menus / measurements.
- INFO field shows information / instructions to navigate.
- USER field shows current user name.
- MODE field shows whether the instrument is working in Auto, Manual or Run mode.
- STATUS field shows current working status of the instrument.
- FUNCTIONAL KEYS ASSIGNMENT FIELD shows the functions allotted to the functional keys.
- FUNCTION KEYS select functions.
- OPERATIONAL KEYS are meant for operation of the instrument.
WORK
INSTRUCTION PROCEDURE:
·
Check the tightness
of all cable terminations & tube fittings.
·
Ensure healthiness of LCD & sensor.
·
If transmitter showing error/bad reading,
sensor to be checked.
·
Before switching off the power supply, ensure
transmitter display is working normally.
·
If transmitter showing error/bad reading.
·
Switch on the power supply
· Reading in transmitter display should match with DCS.
THE WORKING PRINCIPLE:
·
The colorimetric method
is based on the reaction of reagents with the sample to produce a color. This
color is measured and compared to a standard color, which proportionally
defines its concentration.
·
The color compound,
which is the reacted mixture of the sample with the reagents, absorbs energy at
a specific wavelength in accordance to Beer-Lambert Law.
4.3.1
DETAILS OF HAZARD INVOLVED:
·
Electric Shock.
·
Fire
·
Injury due to improper
handling of Regents
·
Skin burning
·
These warnings identify
areas of potential hazard, which could result in personal injury, or loss of
life.
4.3.2
SAFETY PRECAUTION REQUIRED:
·
The Thermo Scientific 2295
Phosphate Analyzer has been designed with careful attention to safety. It
complies with formal safety design rules established by international
regulatory agencies. Observe all safety warnings marked on the instrument.
·
To ensure that the
protection provided by this equipment is not impaired, do not use or install
this equipment in any manner other than that which is specified in this manual.
·
Do not use this product
for any purpose other than described in this manual.
·
Disconnect power before
servicing and prior to making any terminal connections within the Analyzer
enclosure.
·
Do not operate with the electronics enclosure
door open.
·
Ensure proper
ventilation at the work place
·
Ensure use of proper
PPE’s
·
Keep the fire
extinguishers (CO2 gas / DCP) ready within reachable limit for any exigency at
the vicinity of work place.
·
Ensure proper Earthing
& proper illumination
·
Ensure proper range of
multi meter & test probe for different measurement.
·
Insulated general tools
to be used.
·
Ensure good
housekeeping of maintenance site.
·
Permit Type: Permit to
Work Order
· Permit Approving Authority: Shift in-charge (Main Control Room).
SPARES & CONSUMABLES:
·
The Phosphate analysis
requires two reagents.
·
The reagent usage depends on the frequency of
analysis. Typically 0.23 mL of each reagent is used per analysis cycle.
·
Based on 4 analyses per
hour, the 1 Liter reagents should last approximately 45 days.
Parts and Accessories
Catalog Number
Description |
|
Maintenance and Service Items |
|
2295MK |
Maintenance Kit |
2230PA |
Pump Head Tubing Covers (3
each) |
2230PS |
Power Supply Assy |
2230FS |
Power Fuses |
|
|
2295TK |
Main Tubing Kit |
2230BC |
Empty Auto Cal Btl (Validation Sample Bottle) |
2230SF |
Fittings, Sample In, Sample Out and Drain |
2230RC |
Reagent and Standard Bottle Cap - includes fittings |
Reagents and
Standards |
|
2295REC |
Reagent |
229510 |
10 ppm Calibration Standard (For calibration range 0.2-20 ppm) |
229530 |
30 ppm Calibration Standard (For calibration range 20-40 ppm) |
229550 |
50 ppm Calibration Standard (For calibration range 40-50 ppm) |
5
Maintenance Conduct
regular maintenance Schedule Frequency to ensure the analyzer performs properly. |
|
Every week |
Inspect in / out tubes, reagent bottle, and all tubes for
fouling. |
Every month |
Inspect tubing and ensure that it is
not blocked. |
Every 1.5 months |
Replace reagent if necessary every
45 days. |
Every 12 months |
Replace pump
tubing. Use factory supplied maintenance kit. Inspect for any fouling of the reaction cell. |
ISOLATION
PROCEDURE:
·
To put the System in Stand-by mode, make sure to press the STOP key and
check the DISPLAY field next to F4 to show MANUAL RESTART. If AUTO RESTART is
selected, you need to set the mode to Manual:
·
MENU ➔
SETUP ➔
+ ANALYSIS PARAMETERS ➔ MODE
·
The system will switch
to Stand-By after 60 minutes of no keypad activity. In Stand-By, the analyser
will rotate the pumps periodically to avoid pinched tubing from sealing.
·
To avoid clogging,
flush the entire system with distilled water before shutdown. Disconnect and
empty out the reagents bottles and flush the lines using FLUIDICS PRIME menu.
Release the pressure on the peristaltic pump tubing.
· Use the table below to diagnose and correct analyzer problems.
T&P INCLUDING SPECIAL ONES
·
Multi meter
·
General tools.
·
Tissue paper.
·
DM Water
CALIBRATION:
·
If the result is
outside the user specified tolerance (default 5%), the analyzer will warn the
user to avoid storing the wrong calibration data. The user may then ACCEPT or
REJECT the calibration result.
·
Note user accepted manual calibration will
become the REFERENCE calibration data.
·
The analyzer will use
the REFERENCE calibration data when checking for the acceptable
· Tolerance limits in BOTH manual and auto Cal cycles. gee of using this function over the Run Analysis function is that prior knowledge of the concentration allows for validating the accuracy of the analyzer, without affecting the data reporting to your control center or device.
In Auto-Calibration Mode:
·
The system will run a
second calibration cycle if the first calibration result was outside the user
specified tolerance.
·
If the second calibration result is still
outside the specified tolerance limits,
·
The analyzer will continue
to use the previously stored calibration and will flash the “CHK CAL” message
in the Info section of the LCD screen.
· The “CHK CAL” message can be cleared by resetting the fault alarm under VIEW DATA AND ALARMS ➔ REVIEW/RESET ALARMS or by running a manual calibration.
·
The result of the
calibration will be displayed as follows:
·
STANDARD : 10 ppm
***depends on the range selected
·
ABSORBANCE : 0.19
·
SLOPE : +055.00
·
INTERCEPT : -0.4
·
Running a validation is
identical to running an analysis: Although the results are printed they are not stored, alarm conditions
are not tested, and the analog outputs are not updated.
·
It is highly
recommended to allow the Orion 2295 Phosphate Analyser to complete its
measurement cycle rather than using the STOP key.
·
If the STOP key is
pressed before a measurement cycle is complete, it is highly recommended to
perform both the AUTO PRIME ALL and CLEAN sequences.
Cleaning
·
Automatic flushing of
the reaction cell may be programmed. The default setup for the analyser uses
the sample as the cleaning solution. A cleaning cycle can be performed by
pressing:
·
STOP ➔
MENU ➔
MEASUREMENT ➔
CLEAN
·
This cycle helps
maintain clean reaction cell.
NOTE: To
avoid damage and contamination to the reaction cell, use only cleaning
solutions recommended by the factory.
Post Wash:
·
Automatic post wash
cycles are typically part of every measurement cycle. At the end of the
analysis, the post wash cycle evacuates the reacted sample from the cell and
cleans it to minimize coating and residue deposits in the cell. The post wash
cycle settings can be accessed from the ANALYSIS PARAMETERS menu:
MENU ➔
SETUP ➔
(+) ANALYSIS PARAMETERS ➔ (+)
Please consult
factory before modifying any of the parameters.
Priming:
·
Priming of the
instrument lines may be done as auto (all the solutions) or manually by
selecting the specific reagent or standard. Priming fills the tubing, purges
the lines of air, and prepares the analyzer for measurement mode.
·
The PRIME VOLUME in mL
applies to all the listed solutions in the priming menu.
· The Prime Volume may be specified as a (-) number; this allows the user to pump back the reagents inside the tubing back into the reagents containers allowing for safer service when replacement of the tubing is needed. User may select to pull back all or any one of the reagents or standard.
CHECKS BEFORE RELEASE OF PERMIT
·
All T & P’s and
manpower are removed.
·
Clean the maintenance
area properly.
·
Normalize all isolation & Return PTW
HANDLING OF CONCESSIONS OR DEVIATIONS:
·
In the Event of Any
Changes or Deviation from Procedure are required due to Urgent business requirements
or system/Equipment approval should be obtained from Head O&M.
·
The report for the
concerned task should clearly Captured the details of deviations and reason for
the same.
5
REFERENCES:
THERMO-SCIENTIFIC ORION PHOSPHATE (2295)
Post a Comment